MedRadsQC © – Radiography Quality Control system for Medical Radiation Instruments

by Herdi and Zahri, Universiti Sains Malaysia

Contact: Alamat emel ini dilindungi dari Spambot. Anda perlu hidupkan JavaScript untuk melihatnya.

Facility & Equipment Information

1. Generator Performance Test

1.1 Repeatability of Radiation Output

Optimum Standard: (Coefficient of Variation (CoV): ≤ 5%)

Exposure FactorMeasured Readings
kVmAskVDose
Mean
Standard Deviation (SD)
CoV (%)
Result: Pending

1.2 Accuracy of kVp

Set kV < 100kV (Maximum Deviation: ± 5kV)

Set kVMeasuredDev (kV)

Set kV ≥ 100kV (Maximum Deviation: ± 5%)

Set kVMeasuredDev (%)
Result: Pending

1.3 Accuracy of Exposure Timer

Timer < 0.1sec (Maximum Deviation: ≤ ±20%)

Set secMeasuredDev (%)

Timer ≥ 0.1sec (Maximum Deviation: ≤ ±10%)

Set secMeasuredDev (%)
Result: Pending

1.4 Coefficient of Linearity

Optimum Standard: (Coefficient of Linearity: ≤ 10%)

Set mASet mAsMeasured DoseDose / mAs
Coefficient of Linearity:
Result: Pending

2. X-ray Beam Filtration (HVL measurement)

kV SetMeasured HVL (mm)Min Required (mm)Finding
Reference Minimum HVL
kVp SetMin. HVL (mm)
400.4
501.2
601.3
701.5
802.3
902.5
1002.7
1103.0
1203.2
1303.5
1403.8
1504.1
Result: Pending

3. X-ray Beam Limitation

3.1 Collimation Light Field Illumination

Limit: (Illumination: ≥ 160 lux)

R1 (lux)R2 (lux)R3 (lux)R4 (lux)Average (lux)
Result: Pending

3.2 X-ray Beam / Light Field Alignment

Limit: (Maximum misalignment ≤ 1% of FID on any one side)

Axis PlaneFID (cm)Measured Deviation (cm)Max Allowed (cm)
Parallel
Perpendicular
Result: Pending

4. Image Quality

4.1 Resolution

Limit: (Pattern group visible cannot less than 2 groups from baseline)

Exposure Factor (kV / mAs)Baseline Group No.Comparison Group No.
Result: Pending

4.2 Contrast

Limit: (No. of visible disc cannot less than 2 groups of disc from baseline)

ParameterBaseline Disc No.Comparison Disc No.
High Contrast (0.5mm disc)
Low Contrast (11mm disc)
Result: Pending

5. Focal Spot Size Measurement

Limit: (Fperp or Fparallel ≤ 2.0 x Fnom)

ParameterValue
Star pattern degree (θ)
Tube Focal Spot size Nominal (Fnom)
Maximum limit (2.0 x Fnom)
Star pattern diameter (mm)
Star pattern image diameter (mm)
Distance of blur perpendicular (mm)
Distance of blur parallel (mm)
Measured Fperp (mm)
Measured Fparallel (mm)
Result: Pending

6- Leakage Radiation

(Leakage Radiation ≤ 10mR in an hour)

Exposure Factor
kV
mA
mAs
sec
Leakage Technique Factor
kV
mA
Position Anode Front Cathode Back
Measured Readings (mR/hour)
RC
CF
True Readings (mR/hour)
Normalised Readings (mR in an hour)
Maximum Reading
Result: Pending

7- Scattered Radiation

(Scattered Radiation ≤ 10mR per week)

kV mA sec mAs Patient per week Exposure per patient
Position
Measured Readings (µR/hour)
RC
CF
True Readings (mR/hour)
Weekly dose (mR per week)
Maximum Reading
Result: Pending

SUMMARY

No. Parameter Tested Pass/Fail
1.1Repeatability of Radiation OutputPending
1.2Accuracy of kVpPending
1.3Accuracy of Exposure TimerPending
1.4Coefficient of LinearityPending
2.0X-ray Beam Filtration (HVL measurement)Pending
3.1Collimation Light Field IlluminationPending
3.2X-ray Beam / Light Field AlignmentPending
4.1Image Quality - ResolutionPending
4.2Image Quality - ContrastPending
5.0Focal Spot Size MeasurementPending
6.0Leakage RadiationPending
7.0Scattered RadiationPending