MedRadsQC
MedRadsQC © – Radiography Quality Control system for Medical Radiation Instruments
by Herdi and Zahri, Universiti Sains Malaysia
Contact:
Facility & Equipment Information
1. Generator Performance Test
1.1 Repeatability of Radiation Output
Optimum Standard: (Coefficient of Variation (CoV): ≤ 5%)
| Exposure Factor | Measured Readings | ||
|---|---|---|---|
| kV | mAs | kV | Dose |
| Mean | |||
| Standard Deviation (SD) | |||
| CoV (%) | |||
Result: Pending
1.2 Accuracy of kVp
Set kV < 100kV (Maximum Deviation: ± 5kV)
| Set kV | Measured | Dev (kV) |
|---|---|---|
Set kV ≥ 100kV (Maximum Deviation: ± 5%)
| Set kV | Measured | Dev (%) |
|---|---|---|
Result: Pending
1.3 Accuracy of Exposure Timer
Timer < 0.1sec (Maximum Deviation: ≤ ±20%)
| Set sec | Measured | Dev (%) |
|---|---|---|
Timer ≥ 0.1sec (Maximum Deviation: ≤ ±10%)
| Set sec | Measured | Dev (%) |
|---|---|---|
Result: Pending
1.4 Coefficient of Linearity
Optimum Standard: (Coefficient of Linearity: ≤ 10%)
| Set mA | Set mAs | Measured Dose | Dose / mAs |
|---|---|---|---|
| Coefficient of Linearity: | |||
Result: Pending
2. X-ray Beam Filtration (HVL measurement)
| kV Set | Measured HVL (mm) | Min Required (mm) | Finding |
|---|---|---|---|
| Reference Minimum HVL | |
|---|---|
| kVp Set | Min. HVL (mm) |
| 40 | 0.4 |
| 50 | 1.2 |
| 60 | 1.3 |
| 70 | 1.5 |
| 80 | 2.3 |
| 90 | 2.5 |
| 100 | 2.7 |
| 110 | 3.0 |
| 120 | 3.2 |
| 130 | 3.5 |
| 140 | 3.8 |
| 150 | 4.1 |
Result: Pending
3. X-ray Beam Limitation
3.1 Collimation Light Field Illumination
Limit: (Illumination: ≥ 160 lux)
| R1 (lux) | R2 (lux) | R3 (lux) | R4 (lux) | Average (lux) |
|---|---|---|---|---|
Result: Pending
3.2 X-ray Beam / Light Field Alignment
Limit: (Maximum misalignment ≤ 1% of FID on any one side)
| Axis Plane | FID (cm) | Measured Deviation (cm) | Max Allowed (cm) |
|---|---|---|---|
| Parallel | |||
| Perpendicular |
Result: Pending
4. Image Quality
4.1 Resolution
Limit: (Pattern group visible cannot less than 2 groups from baseline)
| Exposure Factor (kV / mAs) | Baseline Group No. | Comparison Group No. |
|---|---|---|
Result: Pending
4.2 Contrast
Limit: (No. of visible disc cannot less than 2 groups of disc from baseline)
| Parameter | Baseline Disc No. | Comparison Disc No. |
|---|---|---|
| High Contrast (0.5mm disc) | ||
| Low Contrast (11mm disc) |
Result: Pending
5. Focal Spot Size Measurement
Limit: (Fperp or Fparallel ≤ 2.0 x Fnom)
| Parameter | Value |
|---|---|
| Star pattern degree (θ) | |
| Tube Focal Spot size Nominal (Fnom) | |
| Maximum limit (2.0 x Fnom) | |
| Star pattern diameter (mm) | |
| Star pattern image diameter (mm) | |
| Distance of blur perpendicular (mm) | |
| Distance of blur parallel (mm) | |
| Measured Fperp (mm) | |
| Measured Fparallel (mm) |
Result: Pending
6- Leakage Radiation
(Leakage Radiation ≤ 10mR in an hour)
| Exposure Factor | |
|---|---|
| kV | |
| mA | |
| mAs | |
| sec | |
| Leakage Technique Factor | |
|---|---|
| kV | |
| mA | |
| Position | Anode | Front | Cathode | Back |
|---|---|---|---|---|
| Measured Readings (mR/hour) | ||||
| RC | ||||
| CF | ||||
| True Readings (mR/hour) | ||||
| Normalised Readings (mR in an hour) | ||||
| Maximum Reading | ||||
Result: Pending
7- Scattered Radiation
(Scattered Radiation ≤ 10mR per week)
| kV | mA | sec | mAs | Patient per week | Exposure per patient |
|---|---|---|---|---|---|
| Position | |||||||
|---|---|---|---|---|---|---|---|
| Measured Readings (µR/hour) | |||||||
| RC | |||||||
| CF | |||||||
| True Readings (mR/hour) | |||||||
| Weekly dose (mR per week) | |||||||
| Maximum Reading | |||||||
Result: Pending
SUMMARY
| No. | Parameter Tested | Pass/Fail |
|---|---|---|
| 1.1 | Repeatability of Radiation Output | Pending |
| 1.2 | Accuracy of kVp | Pending |
| 1.3 | Accuracy of Exposure Timer | Pending |
| 1.4 | Coefficient of Linearity | Pending |
| 2.0 | X-ray Beam Filtration (HVL measurement) | Pending |
| 3.1 | Collimation Light Field Illumination | Pending |
| 3.2 | X-ray Beam / Light Field Alignment | Pending |
| 4.1 | Image Quality - Resolution | Pending |
| 4.2 | Image Quality - Contrast | Pending |
| 5.0 | Focal Spot Size Measurement | Pending |
| 6.0 | Leakage Radiation | Pending |
| 7.0 | Scattered Radiation | Pending |